5 Channel Calibrator
The 5 Channel Calibrator is designed to interface CEM analyzers with five individual calibration gasses. The gasses can be controlled manually or remotely to flow through one of two calibration ports. Sub-Rack designed for installation into a standard 19" instrument rack. Flow-meter with stainless steel needle valve provides indication and control of calibration gas flow rate from 1 to 10 lpm. Auto Mode allows a data acquisition system equipped with a dry contact closure to control the calibration points remotely. A 12VDC output signal is provided for use by the control system via an Amphenol connector located on the rear of the 5 Channel Calibrator.
Features
Flow meter with stainless steel needle valve provides indication and control of calibration gas flow rate from 1 to 10 lpm.
AUTO Mode allows a data acquisition system equipped with a dry contact closure to control the calibration points remotely.
A 12 VDC output signal is provided for use by the control system via an Amphenol connector located on the rear of the calibrator. Sub-Rack Design for installation into a standard 19" instrument rack.
Part Number
This item may also be rented through CleanAir Rentals